MILWAUKEE ELECTRIC TOOL CORPORATION – TEST & MEASUREMENT PRODUCT
REPAIR SERVICE INSTRUCTIONS — CALIBRATION
2266-20 Laser TEMP-GUN™ Thermometer
Environmental
Condition
REPAIR SERVICE
INSTRUCTIONS
Calibration
MILWAUKEE TOOL
13135 WEST LISBON ROAD BROOKFIELD, WISCONSIN 53005-2550 USA
58-92-2266d3
Calibration and verification are performed at an ambient temperature of 23°C ~ 25°C
and ambient relative humidity <55% in a controlled room. Each thermometer should
be uniquely identified as to the room and/or area in which it is used or tested.
www.milwaukeetool.com03/2014
Seitenansicht 0
1 2 3 4 5 6 7 8 9

Inhaltsverzeichnis

Seite 1 - Calibration

MILWAUKEE ELECTRIC TOOL CORPORATION – TEST & MEASUREMENT PRODUCT REPAIR SERVICE INSTRUCTIONS — CALIBRATION 2266-20 Laser TEMP-GUN™ Thermometer E

Seite 2 - TABLE OF CONTENTS Page

i TABLE OF CONTENTS PageIntroduction 11 noitamrofnI ytefaS dna snoituacerPSymbols 1Safety 2Specifications 3 General Specifications 3Basic Ma

Seite 3 - Introduction

1 Introduction WARNINGTo avoid shock or injury, do not perform the verification tests or calibration procedures described in the manual unless you are

Seite 4 - Specific precautions

2 SAFETY Review the following safety precautions to avoid injury and prevent damage to this product or products connected to it. To avoid potential h

Seite 5 - SPECIFICATIONS

3 SPECIFICATIONS All specifications are warranted unless noted typical and apply to the 2266-20 Laser TEMP-GUN™Thermometer.Stated accuracies are at 2

Seite 6

4 Basic Maintenance WARNINGTo reduce the risk of injury, always remove the batteries from the tool before performing any maintenance. Neve

Seite 7

51.0 TESTING PERSONNEL QUALIFICATION: 1.1 Testing personnel are required to be knowledgeable of the procedures in this manual.2.0 EQUIPMENT USE:

Seite 8

64.0 PRODUCTION CALIBRATION AND TESTING PROCEDURE: 4.1 IR Software Calibration Check using Black Body Source (Calibrator 9133 or 4181) 4.1.

Seite 9

7 4.1.6 You can skip any calibration points by pressing “ALARM” button until the unit will exit and turn off. 4.1.7 Once exit, the uni